Common use of Repeatability Clause in Contracts

Repeatability. The repeatability, defined as 2.5 times the standard deviation of ten repetitive responses to a given calibration or span gas, must be no greater than ± 1% of full scale concentration for each range used above 155 ppm (or ppm C) or ± 2% of each range used below 155 ppm (or ppm C).

Appears in 1 contract

Sources: Agreement Concerning the Adoption of Uniform Conditions of Approval and Reciprocal Recognition of Approval for Motor Vehicle Equipment and Parts

Repeatability. The repeatability, defined as 2.5 times the standard deviation of ten repetitive responses to a given calibration or span gas, must be no greater than ± 1% 1 per cent of full scale concentration for each range used above 155 ppm (or ppm C) or ± 2% 2 per cent of each range used below 155 ppm (or ppm C).