Door. 2.1 Replace chamber door seal. 2.2 Test door locking mechanism (Examine ratchet set, replace diaphragm). 2.3 Lubricate nose-piece and spindle, report to authorities if threads are worn and not safe. 2.4 Check wearing strips and fingers, report to authorities if replacement needed. 2.5 Test clutch on sliding door, adjust or repair if necessary. 2.6 Check chains and hinges if fitted, repair or replace if necessary. 2.7 Visually check door for cracks. 2.8 Visually check safety flap operation where fitted.
Appears in 1 contract
Sources: Contract for the Supply, Maintenance and Repair of Autoclave Installations
Door. 2.1 Replace chamber door sealseal and replace if necessary.
2.2 Test door locking mechanism (Examine ratchet set, replace diaphragm)mechanism.
2.3 Lubricate nose-piece and spindle, report to authorities Authorities if threads are worn and not safe.
2.4 Check wearing strips and fingers, report to authorities Authorities if replacement needed.
2.5 Test clutch on sliding door, adjust or repair if necessary.
2.6 Check chains and hinges if fitted, repair or replace if necessary.
2.7 Visually check door for cracks.
2.8 Visually check Check operation of safety flap operation where flaps is fitted.