Cooling. SPECIFICATION: Rate: Ambient -40(degrees)C in 30 minutes Rate: 150(degrees)C to -40(degrees)C in 30 minutes Accuracy: + or (-) 5(degrees)C or 5% of setpoint, whichever is greater VALIDATION PROCEDURE: Make sure a LN cryo tank, with sufficient capacity has been connected before you run the above "Heating and Cooling test". After the run, check the runfile for time and temperature readings in the cooling ranges. Verify specifications are met. Attach runfile to installation validation documentation. RESULT: DIVERSITY REAGENT DELIVERY ACCURACY/PRECISION (AUTOSAMPLER) SPECIFICATION: Range: minimum 50 (nanoliters) to 3 ml in 5 (nanoliter) increments Accuracy: + or (-) 15 (nanoliters) (<200(nanoliters)): + or (-) 5% (>200 (nanoliters)) VALIDATION PROCEDURE: Run the preprogrammed "Solvent/Reagent Delivery Test" utilizing water or DMF in the AutoSampler. Reagent and Solvent locations, and a cassette with 48 fully prepared vials. Measure the volume in the collection vials which correspond to AutoSampler deliveries using gravemetric determination, or transfer to a graduated cylinder. All deliveries tested should be within the accuracy range. RESULT: (Argonaut Installation Engineer to enter result here and attach supporting documentation)