CTDI100 definition

CTDI100 means the accumulated multiple scan dose at the center of a 100-mm scan and underestimates the accumulated dose for longer scan lengths. It is thus smaller than the equilibrium dose. The CTDI100, requires integration of the radiation dose profile from a single axial scan over specific integration limits. In the case of CTDI100, the integration limits are +50 mm, which corresponds to the 100-mm length of the commercially available "pencil" ionization chamber. CTDI100 is acquired using a 100-mm long, 3-cc active volume CT "pencil" ionization chamber and one of the two standard CTDI acrylic phantoms (16 and 32 cm diameter) and a stationary patient table. The equation is:
CTDI100 means the accumulated multiple scan dose at the center of a one hundred millimeter (100 mm) scan and underestimates the accumulated dose for longer scan lengths. It is thus smaller than the equilibrium dose. The CTDI100, requires integration of the radiation dose profile from a single axial scan over specific integration limits. In the case of CTDI100, the integration limits are + fifty millimeters (50 mm), which corresponds to the one hundred millimeters (100 mm) length of the commercially available “pencil” ionization chamber. CTDI100 is acquired
CTDI100 means the accumulated multiple scan dose at the center of a 100-mm

Examples of CTDI100 in a sentence

  • By comparison, this dose is up to 2-3 times lower than multi-slice CT scans of the breast, which results in a CTDI100 ~12-18 mGy glandular dose/scan.