PRECISION. The precision studies were performed using three lots of Cascade Abrazo PT-C Level 1 and Level 2 controls and one lot of Cascade Abrazo PT-C cards. The studies were performed by multi non-laboratorian operators (POC) at three sites across 6 Cascade Abrazo analyzers. Each operator performed 2 runs per day, 2 tests per run on each lot of Abrazo PT-C test controls over a period of 20 days. *Precision studies were performed according to EP5-A2.6
Appears in 2 contracts
Sources: End User Agreement, End User Agreement
PRECISION. The precision studies were performed using three lots of Cascade Abrazo PT-C WB Level 1 and Level 2 controls and one lot of Cascade Abrazo PT-C WB cards. The studies were performed by multi multiple non-laboratorian operators (POC) at three sites across 6 Cascade Abrazo analyzers. Each operator performed 2 runs per day, 2 tests per run on each lot of Abrazo PT-C WB test controls over a period of 20 days. *Precision studies were performed according to EP5-A2.6A2.6 References
Appears in 2 contracts
Sources: End User Agreement, End User Agreement