Parameter Measured Clause Samples

Parameter Measured. Specific Conductivity‌
Parameter Measured. Velocity‌ Simple diagnostic procedures are provided to verify system operation. Beam Check requires an external computer. The Auto QC Test procedure requires only a few minutes and can be performed in the field from the keypad interface. These should be performed before each data run. Beam Check is a diagnostic program that is used to verify FlowTracker performance. This is the same diagnostic program used at SonTek; it provides you with a powerful tool for understanding and verifying system performance. We recommend you become familiar with this software and use it on a regular basis. To run Beam Check: 1. Hold the FlowTracker in a bucket of water (or a natural environment) such that the probe is submerged and there is a boundary (surface, side, or bottom) within view. a. Ideally, the boundary should be placed 20-30 cm (8-12 in) from the probe. b. You may need to add a small amount of fine dirt or other seeding material and stir the bucket well for good test conditions. Regular tap water usually does not have enough scatterers (seeding) for a valid test. 2. Connect the FlowTracker to the PC and turn the system on. a. Run the FlowTracker software (click Start | Programs | SonTek Software | FlowTracker).