LEAK TEST CYCLE Clause Samples

LEAK TEST CYCLE. NOTE: S7 solenoid valve is used to control the waste water temperature below 140°F. S7 will turn on any time it is necessary to cool the waste temperature. NOTE: Components shown in parentheses ( ) denote components used only on double door sterilizers. 1. With the display showing the Main Menu, press the Leak Test Cycle button.‌‌ 2. Connect absolute gauge to chamber. Keep ball valve closed at this time. Close door(s) if neces- sary. 2.5.1 Activate Seal(s)