Coriolis Meters. In accordance with American Gas Association Report No. 11, and any subsequent amendments, revisions, or modifications, thereof. Notwithstanding anything contained in this Exhibit to the contrary, Gatherer shall not be required to replace or make any alterations to its measuring equipment as a result of any subsequent amendments, revisions, or modifications of the American Gas Association Reports cited in Subparagraphs (1) through (6) of this Section 2(a), unless the Parties mutually agree to such replacement or alteration.
Coriolis Meters. Coriolis type mass flowmeters shall generally be preferred for critical applications requiring highly accurate and reliable measurements of mass flow and/or density. The mass flow reading shall be insensitive to changes in temperature, pressure, flow profile and fluid viscosity. The mass flow reading shall also be unaffected by the presence of small volumes of air or gas in the flowing stream. The transmitter shall be of the “smart” type with modular microprocessor based electronics suitable for installation in the ambient conditions described in Sections 12.0 and 13.0 of this specification. The transmitter shall have two analog 4-20 mA outputs configurable for flow rate, density, or temperature. For further information, refer to Project Specification 5578-E2-SP-IE-00001, Flow Instruments.